AMETEK i-iX Series II User Manual
Page 169
User Manual
California Instruments
i Series II / iX Series II / iM Series II
169
9.1.1.5.3 ABNORMAL TEST
From the DO160 MENU Scroll to the ABNORMAL AC entry using the up and down cursor keys.
Press the ENTER key to select the ABNORMAL screens. The screen will appear as shown in
Figure 9-9.
Figure 9-9: Abnormal Screen
The ABNORMAL SCREEN has the following tests:
1
VOLT MAX
2
VOLT MIN
3
VOLT SURG
4
VOLT DROP
5
FREQ TRANSIENTS
The above test can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test.
VOLT MAX
Standard/Group
RTCA
Group1
Group2
Group3
Voltage
1
97
97
104
97
97
3
98.5
98.5
105.5
98.5
98.5
Frequency
400
400
370
360
360
Table 9-10: Abnormal Voltage Minimum
Standard/Group
RTCA
Group1
Group2
Group3
Voltage
1
134
134
122
134
134
3
120.5
132.5
120.5
132.5
132.5
Frequency
400
400
430
650
800
Table 9-11: Abnormal Voltage Maximum
This test will set the voltage and frequency to levels defined by Figure 9-10 for 5 minutes. The
test will be repeated for Group1 only as indicated in Table 9-10 for voltage and Table 9-11 for
frequency. All Groups will repeat the test using Table 9-10 for the voltage setting and Table 9-
10 or Table 9-11for the frequency setting. The
←
key (backspace) will terminate the test at any
time.