AMETEK i-iX Series II User Manual

Page 255

background image

User Manual

California Instruments

i Series II / iX Series II / iM Series II

255

Number

Message String

Cause

Remedy

23

"ALC or Impedance
must be off"

Conflict between ALC
and programmable
impedance mode.

Turn off ALC to use programmable
impedance. Turn off programmable
impedance to use ALC.

24

“Output Relay must be
closed”

Attempt to change
voltage range while
output relay is closed.

Open output relay first, then change
range.

25

“Overvoltage
Protection Trip”

Over voltage trip limit
exceeded.

Check for inductive kickbacks or
overshoot on eut.

26

“Peak Current
Protection Trip”

Peak current exceeds
trip limit.

Reduce load or increase trip level.

29

“DC component
exceed limit”

Waveform selected
has too much DC
content to be
supported on AC
coupled voltage
range.

Use AC waveform without DC offset
only or select DC coupled voltage
range. (Does not apply to iM/i/iX
Series. Applies to MX Series with –
HV models only).

Table 9-22: Error Messages.