beautypg.com

15 switching on uv light for sterilisation cycle – EVCO EVFTFT818P7U Installer manual User Manual

Page 44

background image

EVCO S.p.A.

EVFTFT818 | Installer manual ver. 2.1 | Code 144FTFT818E214

page 44 of 94

Operate as indicated to cut-off pre-cooling:

4.

Press and hold the START/STOP key 3 s.

Parameter r120 establishes the work set-point during pre-cooling.

When the cabinet temperature reaches that established with parameter r12, pre-cooling continues and the buzzer is

activated for 2 s.

6.14

Management of the test regarding correct insertion of

the needle probe

If the needle probe is enabled, i.e. the parameter P3 is set at values different to 0; the temperature-controlled cycles

are preceded by a test on two phases for the verification of the correct insertion of the needle probe.

The second phase is only performed if the first is not completed successfully.

The first phase is completed successfully if the "temperature detected by the needle probe - cabinet temperature"

difference is greater than the value established with parameter r17 in at least 3 controls out of 5 (the controls are

performed at 10 s intervals, consider the difference without sign).

The second phase is completed successfully if the "temperature detected by the needle probe - cabinet temperature"

difference is 1°C/1°F higher with respect to the previous control in at least 6 controls out of 8 (the controls are

performed at time intervals corresponding to 1/8 of the time established with parameter r18; consider the difference

without sign).

If the test is completed successfully, the cycle will be started; if the test is not completed successfully, the buzzer will

be activated for 5 s every 10 s and the cycle will be started with timed-control.

To start the temperature-controlled cycle, press the

key or the

key. After 1 min from the signal that the test has

not been completed successfully without having operated, the cycle is started with time control.

If parameter r17 is set at 0, the test will not be carried out (neither first nor second phase).

6.15

Switching on UV light for sterilisation cycle

Operate as follows:

1.

Make sure that parameter u11 is set at 2.

2.

Make sure the device is in the "on" status and that the door is closed, i.e. the door micro switch is not active.

3.

Make sure that the keyboard is not locked and that no procedure is in progress.

4.

Press and release the

key (1), press and release the

key (2) and then press and release the

START/STOP key (3). The device will display the residual time of the UV light switch-on duration and the

cabinet temperature.

The UV light is switched on for the time period established by parameter u6; opening the door i.e. the activation of the

door micro switch cause the light to switch off.