Amplitude calibration for transverters – Elecraft P3 High-Performance Panadapter Manual User Manual
Page 32
32
Select
MENU:Lvl Cal
, tap the
SELECT
knob, and adjust the knob to place the
signal peak at -73 dBm on the display.
Note: the K3’s RX ANT input has about 1
dB less loss than the ANT connector, so
you may want to set the signal level at -72
dBM when using that input.
Amplitude Calibration for
Transverters
You can calibrate a receiving or transmitting
converter, either the K3’s internal K144XV or an
external device connected to the K3.
If you plan to do the above calibration on an HF
frequency, do that first since that calibration will
affect the transverter calibration.
Set up the P3, the signal generator, the
transverter and the K3 in the manner
described above to obtain a signal in the
center of the right-hand side of the display
and with its peak centered vertically.
Select
MENU:XV Gain
, tap the
SELECT
knob, and adjust the knob to select the
desired transverter band: 1-9. (You can
enter separate calibration values for each
transverter band by repeating this procedure
for each band.)
Tap
the
SELECT
knob again, and adjust
the knob to place the signal peak at the
level corresponding to the output of your
signal generator.