beautypg.com

1 failure rates and sff for 701150 (ac 230 v), 6 safety manual – JUMO 701150 14597 safetyM STB/STW - Safety Temperature Limiter and Safety Temperature Monitor Operating Manual User Manual

Page 41

background image

6 Safety Manual

41

2014-05-01

6.8

Safety instrumented parameters related to the temperature monitoring unit

The following parameters were calculated by means of an FMEDA component under the following conditions:
-

Error models corresponding to requirements of DIN EN 61508 for conformity with SIL2 or SIL3

-

Failure rate of components according to the RDF 2000 UTE C 80-810 standard and SN 29500

-

Sensors were combined as a subsystem in the following five variants:

6.8.1

Failure rates and SFF for 701150...23 (AC 230 V)

Important information:
Variants 1 to 4 were evaluated with JUMO probes according to data sheets 901006 and 902006.

For variant 5 no sensor system was taken into account.

In this case, the user selects the sensor system.

Variant

λ

s

[Fit]

λ

dd

[Fit]

λ

dd

[Fit]

SFF

PFH (1/h)

PFD

avg

1

865.21

306.24

32.31

96 %

4.56 e

-9

2.02 e

-4

1a

865.21

306.24

32.31

96 %

1.05 e

-9

4.57 e

-5

2

868.17

303.28

32.31

96 %

1.05 e

-9

4.57 e

-5

3

881.62

326.78

33.62

96 %

1.03 e

-9

4.49 e

-5

4

887.68

343.82

35.52

96 %

1.22 e

-9

5.30 e

-5

5

881.02

313.43

35.57

96 %

1.04 e

-9

4.48 e

-5