Test and pipette conditions

Eppendorf SOP

English (EN)

1.2

Measuring station

1.2.1

Evaporation

Take evaporation protection into account while carrying out the measurement.

With volumes < 50 μl in particular, errors caused by the evaporation of the test
liquid must be taken into consideration.

This can be ensured by using a liquid trap or other equipment to prevent
evaporation.

1.2.2

Measuring station

For electronic measurement data processing, we recommend calibration
software and accessories for your balance (PICASO, order no. 3113 004.001).

(see also EN ISO 8655, Part 6)

1.2.3

Test room

The tests should be carried out in a draft-free room under constant climatic
conditions.

The test room should have a constant temperature between 15 °C and 30 °C and
a constant relative humidity above 50 %.

1.2.4

Temperature differences

Before the test, the device to be tested and test liquid must have stood in the test
room for a sufficient amount of time, at least 2 hours, in order to reach
equilibrium with the storage conditions.

Direct sunlight and other influences which could affect the temperature should
be avoided at all costs.

1.2.5

Test liquid

Distilled or deionized water of "Quality 3 in accordance with ISO 3696",
degassed or at equilibrium with air. The water must be at room temperature.

1.2.6

Operating Manual

Observe the operating instructions for your pipette.