Adjustment
Eppendorf SOP
English (EN)
7.2.5
Research variable
Auxiliary equipment
• Supplied pipette wrench (order no. 3111 501.016)
• Supplied red CAL seal
4. A revolution corresponds to:
5. Pull off the pipette wrench.
6. Move the volume setting ring slightly forwards and backwards so that the
counting and stroke systems engage again.
7. After the adjustment has been successfully completed, lock the adjustment
opening with a red CAL seal.
1. Insert side D of the pipette wrench
vertically into the lateral adjustment
opening of the pipette grip.
2. Move the pipette wrench into the
vertical position.
3. Turn the volume setting ring toward
– or +.
This adjusts the piston stoke of the
pipette. The digital display does not
change in the process.
Volume range
Vol. per revolution
0.1 – 2.5 μL
approx. 0.1 μL
0.5 - 10 μL
approx. 0.5 μL
2 - 20 μL
approx. 1 μL
10 - 100 μL
approx. 5 μL
20 - 200 μL
approx. 10 μL
100 - 1 000 μL
approx. 50 μL
500 - 5 000 μL
approx. 250 μL
1 - 10 mL
approx. 510 μL