beautypg.com

8 research multi-channel, 9 research plus multi-channel, Research multi-channel – Eppendorf SOP User Manual

Page 31: Research plus multi-channel

background image

31

Adjustment

Eppendorf SOP

English (EN)

7.2.8

Research Multi-channel

Auxiliary equipment
• Supplied pipette wrench (order no. 3111 501.016)
• Supplied red ADJ seal

4. A revolution corresponds to:

5. Pull off the pipette wrench.
6. Move the volume setting ring slightly forwards and backwards so that the

counting and stroke systems engage again.

7. After the adjustment has been successfully completed, lock the adjustment

opening with a red ADJ-seal.

7.2.9

Research plus Multi-channel

Carry out the adjustment as described for the Research plus single-channel
pipette (see p. 25).

1. Insert the D side of the pipette

wrench horizontally into the
adjustment opening on the side of
the pipette grip.

2. Move the pipette wrench into the

vertical position.

3. Turn the volume setting ring toward

– or +. This adjusts the piston stoke
of the pipette. The digital display
does not change in the process.

Volume range

Vol./turn

0.5 - 10 μL

approx. 0.5 μL

10 - 300 μL

approx. 5 μL

30 μL

approx. 10 μL

C

B

A

D