8 research multi-channel, 9 research plus multi-channel, Research multi-channel – Eppendorf SOP User Manual
Page 31: Research plus multi-channel
31
Adjustment
Eppendorf SOP
English (EN)
7.2.8
Research Multi-channel
Auxiliary equipment
• Supplied pipette wrench (order no. 3111 501.016)
• Supplied red ADJ seal
4. A revolution corresponds to:
5. Pull off the pipette wrench.
6. Move the volume setting ring slightly forwards and backwards so that the
counting and stroke systems engage again.
7. After the adjustment has been successfully completed, lock the adjustment
opening with a red ADJ-seal.
7.2.9
Research plus Multi-channel
Carry out the adjustment as described for the Research plus single-channel
pipette (see p. 25).
1. Insert the D side of the pipette
wrench horizontally into the
adjustment opening on the side of
the pipette grip.
2. Move the pipette wrench into the
vertical position.
3. Turn the volume setting ring toward
– or +. This adjusts the piston stoke
of the pipette. The digital display
does not change in the process.
Volume range
Vol./turn
0.5 - 10 μL
approx. 0.5 μL
10 - 300 μL
approx. 5 μL
30 μL
approx. 10 μL