Adjustment
Eppendorf SOP
English (EN)

7.2.7

Research plus adjustable volume - change to the factory
adjustment

Auxiliary equipment
• Supplied safety plug tool
• Supplied pin to loosen the safety plug

You can recognize a change to the factory adjustment that was made by the user
of the Research plus from the red safety plug behind the ejector. If the Research
plus has been adjusted and calibrated by Eppendorf AG, this is indicated by a
gray safety plug.

The factory adjustment can be changed with
the corresponding accessories on a Research
plus with variable volume setting.

1. Check whether the adjustment display on

the side is set to “0”.
If the adjustment display is not set to “0”, it
must be set to “0” first using the adjustment
tool. In this case, instead of continuing with
the factory adjustment changes, carry out a
gravimetric test of the Research plus with
the adjustment display set to “0”.