Troubleshooting
Eppendorf SOP
English (EN)
Liquid cannot be
aspirated correctly
or dripping occurs.
• Leakage in the Combitip
advanced.
Replace the Combitip
advanced with a new
Combitip advanced.
• Combitip advanced has
heated up.
Ensure that the
temperature is uniform as
the liquid expands when
heated.
The control button
is jammed, runs
jerky.
• The piston is
contaminated.
Clean the piston and
lubricate slightly.
• The seal is contaminated.
Disassemble the pipette,
clean all seals, replace if
necessary.
• The piston is damaged.
Replace the piston and
piston seal and lubricate
slightly.
• Penetration of solvent
vapors
Unscrew the lower part of
the pipette and aerate the
lower part.
Clean the piston and
lubricate slightly.
Error Cause
Elimination