2 reference 2 adjustable-volume, Reference 2 adjustable-volume – Eppendorf SOP User Manual
Page 42

42
Technical specifications
Eppendorf SOP
English (EN)
9.2.2
Reference 2 adjustable-volume
Reference2 single-channel variable
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Eppendorf AG error limits
Error
Systematic
error
Random error
± %
± µL
± %
± µL
0.1 μL – 2.5 μL
Increment:
0.002 μL
Dark gray
0.1 μL – 10 μL
34 mm
0.1 μL
±48.0 ±0.048 ±12.0 ±0.012
0.25 μL ±12.0 ±0.03 ±6.0 ±0.015
1.25 μL ±2.5 ±0.031 ±1.5 ±0.019
2.5 μL
±1.4 ±0.035 ±0.7 ±0.018
0.5 μL – 10 μL
Increment:
0.01 μL
Medium gray
0.1 μL – 20 μL
40 mm
0.5 μL
±8.0
±0.04
±5.0
±0.0025
1 μL
±2.5 ±0.025 ±1.8 ±0.018
5 μL
±1.5 ±0.075 ±0.8 ±0.04
10 μL
±1.0 ±0,10
±0.4 ±0.04
2 μL – 20 μL
Increment:
0.02 μL
light gray
0.5 μL – 20 μL L
46 mm
2 μL
±5.0
±0,10
±1.5
±0.03
10 μL
±1.2
±0.12
±0.6
±0.06
20 μL
±1.0 ±0.20 ±0.3 ±0.06
2 μL – 20 μL
Increment:
0.02 μL
Yellow
2 μL - 200 μL
53 mm
2 μL
±5.0 ±0.12
±1.5
±0.03
10 μL
±1.2 ±0.12 ±0.6 ±0.06
20 μL
±1.0 ±0.2
±0.3 ±0.06
10 μL – 100 μL
Increment:
0.1 μL
Yellow
2 μL - 200 μL
53 mm
10 μL
±3.0
±0.3
±0.7
±0.07
50 μL
±1.0
±0.5
±0.3 ±0.15
100 μL
±0.8 ±0.8
±0.2
±0.2
20 μL - 200 μL
Increment:
0.2 μL
Yellow
2 μL - 200 μL
53 mm
20 μL
±2.5
±0.5
±0.7
±0.14
100 μL
±1.0
±1.0
±0.3 ±0.3
200 μL
±0.6 ±1.2
±0.2 ±0.4
30 μL – 300 μL
Increment:
0.2 μL
Orange
20 μL – 300 μL
55 mm
30 μL
±2.5
±0.75
±0.7
±0.21
150 μL
±1.0 ±1.5
±0.3
±0.45
300 μL
±0.6
±1.8
±0.2
±0.6
100 μL – 1 000 μL
Increment:
1 μL
Blue
50 μL – 1 000 μL
71 mm
100 μL
±3.0
±3.0
±0.6
±0.6
500 μL
±1.0
±5.0
±0.2 ±1.0
1,000 μL ±0.6 ±6.0
±0.2 ±2.0