2 adjusting adjustable volume pipettes, 1 reference variable, Adjusting adjustable volume pipettes 9.2.1 – Eppendorf SOP User Manual
Page 41: Reference variable
41
Technical specifications
Eppendorf SOP
English (EN)
9.2
Adjusting adjustable volume pipettes
9.2.1
Reference variable
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Error limits
Error
Systematic error Random error
± %
± µL
± %
± µL
0.1 - 2.5 μL
Dark gray
0.1 - 10 μL
34 mm
0.25 μL
±12.0 ±0.03 ±6.0 ±0.015
1.25 μL
±2.5
±0.031 ±1.5 ±0.019
2.5 μL
±1.4
±0.035 ±0.7 ±0.018
0.5 - 10 μL
light gray
0.5 – 20 μL
46 mm
1 μL
±2.5
±0.025 ±1.8 ±0.018
5 μL
±1.5
±0.075 ±0.8 ±0.04
10 μL
±1.0
±0.1
±0.4 ±0.04
2 - 20 μL
light gray
0.5 – 20 μL
46 mm
2 μL
±3.0
±0.06 ±2.0 ±0.04
10 μL
±1.0
±0.1
±0.5 ±0.05
20 μL
±0.8
±0.16 ±0.3 ±0.06
2 - 20 μL
Yellow
2 - 200 μL
53 mm
2 μL
±5.0
±0.1
±1.5 ±0.03
10 μL
±1.2
±0.12 ±0.6 ±0.06
20 μL
±1.0
±0.2
±0.3 ±0.06
10 - 100 μL
Yellow
2 - 200 μL
53 mm
10 μL
±3.0
±0.3
±0.7 ±0.07
50 μL
±1.0
±0.5
±0.3 ±0.15
100 μL
±0.8
±0.8
±0.15 ±0.15
50 – 200 μL
Yellow
2 - 200 μL
53 mm
50 μL
±1.0
±0.5
±0.3 ±0.15
100 μL
±0.9
±0.9
±0.3 ±0.3
200 μL
±0.6
±1.2
±0.2 ±0.4
50 – 250 μL
Blue
50 - 1 000 μL
71 mm
50 μL
±1.4
±0.7
±0.3 ±0.15
100 μL
±1.1
±1.1
±0.3 ±0.3
250 μL
±0.6
±1.5
±0.2 ±0.5
100 – 1 000 μL Blue
50 – 1 000 μL
71 mm
100 μL
±3.0
±3.0
±0.3 ±0.3
500 μL
±1.0
±5.0
±0.2 ±1.0
1,000 μL ±0.6
±6.0
±0.2 ±2.0
500 – 2 500 μL Red
500 – 2 500 μL
115 mm
0.5 mL
±1.5
±7.5
±0.3 ±1.5
1.25 mL ±0.8
±10
±0.2 ±2.5
2.5 mL
±0.6
±15
±0.2 ±5.0