Eppendorf SOP User Manual
Page 29

29
Adjustment
Eppendorf SOP
English (EN)
6. This results in the following approximate volume changes:
The values mentioned are theoretical values and are for orientation purposes
only. The volume changes mentioned apply to each volume setting.
single-channel
Nominal volume
Color code
+½
revolution
+¼
revolution
–¼
revolution
–½
revolution
2.5 μL
Dark gray
0.106 μL
0.053 μL
-0.053 μL
-0.106 μL
10 μL
Medium gray
0.53 μL
0.27 μL
-0.27 μL
-0.53 μL
20 μL
light gray
1.06 μL
0.53 μL
-0.53 μL
-1.06 μL
20 μL
Yellow
1.07 μL
0.54 μL
-0.54 μL
-1.07 μL
100 μL
Yellow
5.4 μL
2.7 μL
-2.7 μL
-5.4 μL
200 μL
Yellow
10.8 μL
5.4 μL
-5.4 μL
-10.8 μL
300 μL
Orange
10.7 μL
5.4 μL
-5.4 μL
-10.7 μL
1,000 μL
Blue
54 μL
27 μL
-27 μL
-54 μL
5 mL
Purple
271 μL
135 μL
-135 μL
-271 μL
10 mL
Turquoise
542 μL
271 μL
-271 μL
–542 μL
multi-channel
Nominal volume
Color code
+½
revolution
+¼
revolution
–¼
revolution
–½
revolution
10 μL
Medium gray
0.53 μL
0.27 μL
-0.27 μL
-0.53 μL
100 μL
Yellow
5.4 μL
2.7 μL
-2.7 μL
-5.4 μL
300 μL
Orange
10.7 μL
5.4 μL
-5.4 μL
-10.7 μL