Technical specifications

Eppendorf SOP

English (EN)

9.6.3

Dispenser (Multipette)

Nominal volume

Error limits

Measurement errors

Systematic

Random

± %

± µL

± %

± µL

0.001 mL

±5.0 %

±0.05 μl

±5.0 %

±0.05 μl

0,002 ml

±5.0 %

±0.1 μl

±5.0 %

±0.1 μl

0,003 ml

±2.5 %

±0.075 μl

±3.5 %

±0,11 μL

0,01 ml

±2.0 %

±0.2 μl

±2.5 %

±0.25 μl

0,02 ml

±1.5 %

±0.3 μl

±2.0 %

±0.4 μl

0,05 ml

±1.0 %

±0.5 μl

±1.5 %

±0.75 μl

0,1 ml

±1.0 %

±1,0 μL

±1.0 %

±1,0 μL

0.2 ml

±1.0 %

±2,0 μL

±1.0 %

±2,0 μL

0.5 ml

±1.0 %

±5,0 μL

±0.6 %

±3,0 μL

1 ml

±1.0 %

±10,0 μL

±0.4 %

±4,0 μL

2 ml

± 0.8 %

±16,0 μL

±0.4 %

±8,0 μL

5 mL

±0.6 %

±30,0 μL

±0.3 %

±15,0 μL

10 mL

±0.5 %

±50,0 μL

±0.3 %

±30,0 μL

25 ml

±0.5 %

±125,0 μL

±0.3 %

±75,0 μL

50 mL

±0.5 %

±250 μL

±0.25 %

±125.0 μL

100 ml

±0.5 %

±500 μL

±0.25 %

±250.0 μL

200 ml

±0.5 %

±1 000 μL

±0.25 %

±500,0 μL