beautypg.com

5 research plus adjustable-volume, Research plus adjustable-volume – Eppendorf SOP User Manual

Page 46

background image

46

Technical specifications
Eppendorf SOP
English (EN)

9.2.5

Research plus adjustable-volume

Research plus adjustable single-channel

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Testing
volume

Eppendorf AG error limits

Error

Systematic

error

Random

error

± %

± µL

± %

± µL

0.1 μL – 2.5 μL
Increment:
0.002 μL

Dark gray
0.1 μL – 10 μL
34 mm

0.1 μL

±48

±0.048 ±12

±0.012

0.25 μL

±12

±0.03 ±6.0

±0.015

1.25 μL

±2.5

±0.031 ±1.5

±0.019

2.5 μL

±1.4

±0.035 ±0.7

±0.018

0.5 μL – 10 μL
Increment:
0.01 μL

Medium gray
0.1 μL – 20 μL
40 mm

0.5 μL

±8.0

±0.04 ±5.0

±0.025

1 μL

±2.5

±0.025 ±1.8

±0.018

5 μL

±1.5

±0.075 ±0.8

±0.04

10 μL

±1.0

±0.1

±0.4

±0.04

2 μL – 20 μL
Increment:
0.02 μL

light gray
0.5 μL – 20 μL L
46 mm

2 μL

±5.0

±0.1

±1.5

±0.03

10 μL

±1.2

±0.12 ±0.6

±0.06

20 μL

±1.0

±0.2

±0.3

±0.06

2 μL – 20 μL
Increment:
0.02 μL

Yellow
2 μL - 200 μL
53 mm

2 μL

±5.0

±0.1

±1.5

±0.03

10 μL

±1.2

±0.12 ±0.6

±0.06

20 μL

±1.0

±0.2

±0.3

±0.06

10 μμL – 100 μL
Increment:
0.1 μL

Yellow
2 μL - 200 μL
53 mm

10 μL

±3.0

±0.3

±1.0

±0.1

50 μL

±1.0

±0.5

±0.3

±0.15

100 μL

±0.8

±0.8

±0.2

±0.2

20 μL - 200 μL
Increment:
0.2 μL

Yellow
2 μL - 200 μL
53 mm

20 μL

±2.5

±0.5

±0.7

±0.14

100 μL

±1.0

±1.0

±0.3

±0.3

200 μL

±0.6

±1.2

±0.2

±0.4

30 μL – 300 μL
Increment:
0.2 μL

Orange
20 μL – 300 μL
55 mm

30 μL

±2.5

±0.75 ±0.7

±0.21

150 μL

±1.0

±1.5

±0.3

±0.45

300 μL

±0.6

±1.8

±0.2

±0.6

100 μL – 1 000 μL
Increment:
1 μL

Blue
50 μL – 1 000 μL
71 mm

100 μL

±3.0

±3.0

±0.6

±0.6

500 μL

±1.0

±5.0

±0.2

±1.0

1,000 μL ±0.6

±6.0

±0.2

±2.0