2 reference 2 fixed, Reference 2 fixed – Eppendorf SOP User Manual
Page 38

38
Technical specifications
Eppendorf SOP
English (EN)
9.1.2
Reference 2 fixed
Fixed-volume Reference2
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Eppendorf AG error limits
Error
Systematic error Random error
± %
± µL
± %
± µL
1 μL
Dark gray
0.1 μL – 10 μL
34 mm
±2.5
±0.025 ±1.8 ±0.018
2 μL
±2.0
±0.04
±1.2
±0.024
5 μL
Medium gray
0.1 μL – 20 μL
40 mm
±1.2
±0.06
±0.6
±0.03
10 μL
±1.0
±0.1
±0.5
±0.05
20 μL
light gray
0.5 μL – 20 μL L
46 mm
±0.8
±0.16
±0.3
±0.06
10 μL
Yellow
2 μL – 200 μL
53 mm
±1.2
±0.12
±0.6
±0.06
20 μL
±1.0
±0.2
±0.3
±0.06
25 μL
±1.0
±0.25
±0.3
±0.075
50 μL
±0.7
±0.35
±0.3
±0.15
100 μL
±0.6
±0.6
±0.2
±0.2
200 μL
±0.6
±1.2
±0.2
±0.4
200 μL
Blue
50 μL – 1 000 μL
71 mm
±0.6
±1.2
±0.2
±0.4
250 μL
±0.6
±1.5
±0.2
±0.5
500 μL
±0.6
±3.0
±0.2
±1.0
1,000 μL
±0.6
±6.0
±0.2
±2.0
Fixed-volume Reference2
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Eppendorf AG error limits
Error
Systematic error Random error
± %
± mL
± %
± mL
2.0 mL
Red
0.5 mL – 2.5 mL
115 mm
±0.6
±0.012
±0.2
±0,004
2.5 mL
±0.6
±0.015
±0.2
±0.005