Adjustment
Eppendorf SOP
English (EN)
7.2.2
Biomaster
Auxiliary equipment
• Supplied pipette wrench (order no. 4910 092.001)
7.2.3
Reference variable
Auxiliary equipment
• Supplied pipette wrench (order no. 4910 092.001)
• Supplied red CAL seal
1. Use the pipette wrench to adjust the
volume display of the pipette, with
the piston stroke unchanged, to the
actual volume value of the
measurement.
2. Pull off the pipette wrench.
3. Set the pipette to the target volume
in the usual manner.
4. Mark the adjustment on the pipette.
1. Insert side B of the pipette wrench
into the adjustment opening in the
lid.
2. Using the pipette wrench, set the
digital display of the pipette, with
unaltered piston stroke, to the actual
volume value of the measurement.
3. Pull off the pipette wrench.
4. Set the pipette to the target volume
in the usual manner.
5. After the adjustment has been
successfully completed, lock the
adjustment opening with a red CAL
seal.