beautypg.com

6 biomaster, 7 varipette, Biomaster – Eppendorf SOP User Manual

Page 47: Varipette

background image

47

Technical specifications

Eppendorf SOP

English (EN)

9.2.6

Biomaster

9.2.7

Varipette

0.5 μL – 5 mL
Increment:
0.005 mL

Purple
0.1 μL – 5 mL
120 mm

0.5 mL

±2.4

±12

±0.6

±3.0

2.5 mL

±1.2

±30

±0.25 ±6.0

5.0 mL

±0.6

±30

±0.15 ±8.0

1 μL – 10 mL
Increment:
0.01 mL

Turquoise
1 μL – 10 mL
165 mm

1.0 mL

±3.0

±30

±0.6

±6.0

5.0 mL

±0.8

±40

±0.2

±10

10.0 mL ±0.6

±60

±0.15 ±15

Model

Pipette tip

Testing
volume

Error limits

Error

Systematic error Random error

± %

± µL

± %

± µL

Biomaster 4830

Mastertip

2 μL

±6.0

±0.12

±4.0

±0.08

3 μL

±5.0

±0.15

±3.0

±0.09

5 μL

±4.0

±0.2

±2.0

±0.1

10 μL

±3.0

±0.3

±1.5

±0.15

20 μL

±2.5

±0.5

±0.8

±0.16

Model

Pipette tip

Testing
volume

Error limits

Error

Systematic error Random error

± %

± mL

± %

± mL

Varipette 4720

Varitip S

2.5 mL

±1.0

±0.025 ±0.2

±0.005

5 mL

±0.4

±0.02

±0.2

±0.01

10 mL

±0.3

±0.03

±0.2

±0.02

Varipette 4720

Varitip P

1 mL

±0.6

±0.006 ±0.2

±0.002

5 mL

±0.5

±0.025 ±0.1

±0.005

10 mL

±0.3

±0.03

±0.1

±0.01

Research plus adjustable single-channel

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Testing
volume

Eppendorf AG error limits

Error

Systematic

error

Random

error

± %

± µL

± %