6 biomaster, 7 varipette, Biomaster – Eppendorf SOP User Manual
Page 47: Varipette
47
Technical specifications
Eppendorf SOP
English (EN)
9.2.6
Biomaster
9.2.7
Varipette
0.5 μL – 5 mL
Increment:
0.005 mL
Purple
0.1 μL – 5 mL
120 mm
0.5 mL
±2.4
±12
±0.6
±3.0
2.5 mL
±1.2
±30
±0.25 ±6.0
5.0 mL
±0.6
±30
±0.15 ±8.0
1 μL – 10 mL
Increment:
0.01 mL
Turquoise
1 μL – 10 mL
165 mm
1.0 mL
±3.0
±30
±0.6
±6.0
5.0 mL
±0.8
±40
±0.2
±10
10.0 mL ±0.6
±60
±0.15 ±15
Model
Pipette tip
Testing
volume
Error limits
Error
Systematic error Random error
± %
± µL
± %
± µL
Biomaster 4830
Mastertip
2 μL
±6.0
±0.12
±4.0
±0.08
3 μL
±5.0
±0.15
±3.0
±0.09
5 μL
±4.0
±0.2
±2.0
±0.1
10 μL
±3.0
±0.3
±1.5
±0.15
20 μL
±2.5
±0.5
±0.8
±0.16
Model
Pipette tip
Testing
volume
Error limits
Error
Systematic error Random error
± %
± mL
± %
± mL
Varipette 4720
Varitip S
2.5 mL
±1.0
±0.025 ±0.2
±0.005
5 mL
±0.4
±0.02
±0.2
±0.01
10 mL
±0.3
±0.03
±0.2
±0.02
Varipette 4720
Varitip P
1 mL
±0.6
±0.006 ±0.2
±0.002
5 mL
±0.5
±0.025 ±0.1
±0.005
10 mL
±0.3
±0.03
±0.1
±0.01
Research plus adjustable single-channel
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Eppendorf AG error limits
Error
Systematic
error
Random
error
± %
± µL
± %
± µL