Technical specifications
Eppendorf SOP
English (EN)
9.6.1
Air-cushion pipettes, fixed and adjustable volume
9.6.2
Positive displacement pipettes (Biomaster)
Nominal volume
Error limits
Measurement errors
Systematic
Random
± %
± µL
± %
± µL
1 μl
±5.0 %
±0.05 μl
±5.0 %
±0.05 μl
2 μl
±4.0 %
±0.08 μl
±2.0 %
±0.04 μl
5 μl
±2.5 %
±0,125 μL
±1.5 %
±0.075 μl
10 μl
±1.2 %
±0,12 μL
± 0.8 %
±0.08 μl
20 μl
±1.0 %
±0.2 μl
±0.5 %
±0.1 μl
50 μl
±1.0 %
±0.5 μl
±0.4 %
±0.2 μl
100 μl
± 0.8 %
±0.8 μl
±0.3 %
±0.3 μl
200 μL
± 0.8 %
±1,6 μL
±0.3 %
±0.6 μl
500 μl
± 0.8 %
±4,0 μL
±0.3 %
±1.5 μl
1 000 μL
± 0.8 %
±8,0 μL
±0.3 %
±3,0 μL
2 000 μL
± 0.8 %
±16,0 μL
±0.3 %
±6,0 μL
5 000 μL
± 0.8 %
±40,0 μL
±0.3 %
±15,0 μL
10 000 μL
±0.6 %
±60,0 μL
±0.3 %
±30,0 μL
For Multi-channel pipettes, the error values are two times the values
specified for single-channel pipettes.
Nominal volume
Error limits
Measurement errors
Systematic
Random
± %
± µL
± %
± µL
5 μl
±2.5 %
±0,13 μL
±1.5 %
±0.08 μl
10 μl
±2.0 %
±0.2 μl
±1.0 %
±0.1 μl
20 μl
±2.0 %
±0.4 μl
± 0.8 %
±0,16 μL
50 μl
±1.4 %
±0,7 μL
±0.6 %
±0.3 μl
100 μl
±1.5 %
±1.5 μl
±0.6 %
±0.6 μl
200 μL
±1.5 %
±3.0 μL
±0.4 %
±0.8 μl
500 μl
±1.2 %
±6.0 μL
±0.4 %
±2.0 μL
1 000 μL
±1.2 %
±12.0 μL
±0.4 %
±4.0 μL