9 technical specifications, 1 fixed-volume pipettes, 1 reference fixed – Eppendorf SOP User Manual
Page 37: Technical specifications 9.1, Fixed-volume pipettes 9.1.1, Reference fixed, 9technical specifications
37
Technical specifications
Eppendorf SOP
English (EN)
9
Technical specifications
The following specifications are valid for the following conditions:
9.1
Fixed-volume pipettes
9.1.1
Reference fixed
Liquid:
Distilled or deionized water
Reference temperature: 20 °C to 25 °C ±0.5 °C
Number of
determinations:
10, in accordance with EN ISO 8655 with original
Eppendorf pipette tips
Technical specifications subject to change!
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Error limits
Error
Systematic error
Random error
± %
± µL
± %
± µL
1 μL
light gray
0.5 - 20 μL L
46 mm
±2.5
±0.025 ±1.8
±0.018
2 μL
±2.0
±0.04
±1.2
±0.024
5 μL
±1.5
±0.075
±0.8
±0.04
10 μL
±1.0
±0.1
±0.5
±0.05
10 μL
Yellow
2 - 200 μL
53 mm
±1.0 ±0.1 ±0.5
±0.05
20 μL
±0.8
±0.16
±0.3
±0.06
25 μL
±0.8
±0.2
±0.3
±0.075
50 μL
±0.7
±0.35
±0.3
±0.15
100 μL
±0.6
±0.6
±0.2
±0.2
200 μL
Blue
50 - 1 000 μL
71 mm
±0.6
±1.2
±0.2
±0.4
250 μL
±0.6
±1.5
±0.2
±0.5
500 μL
±0.6
±3.0
±0.2
±1.0
1,000 μL
±0.6
±6.0
±0.2
±2.0
1 500 μL
Red
500 – 2 500 μL
115 mm
±0.6
±9.0
±0.2 ±3.0
2 000 μL
±0.6
±12
±0.2
±4.0
2,500 μL
±0.6
±15
±0.2
±5.0