4 reference2, Reference2 – Eppendorf SOP User Manual
Page 52

52
Technical specifications
Eppendorf SOP
English (EN)
9.3.4
Reference2
Reference2 Variable multi-channel (8-channel/12-channel)
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Eppendorf AG error limits
Error
Systematic
error
Random
error
± %
± µL
± %
± µL
0.5 μL – 10 μL
Medium gray
0.1 μL – 20 μL
40 mm
0.5 μL
±12.0 ±0.06 ±8.0
±0.04
1
μL
±8.0 ±0.08
±5.0 ±0.05
5
μL
±4.0 ±0.2 ±2.0 ±0.1
10
μL
±2.0 ±0.2 ±1.0 ±0.1
10 μL – 100 μL
Yellow
2 μL - 200 μL
53 mm
10
μL
±3.0 ±0.3 ±2.0 ±0.2
50
μL
±1.0 ±0.5 ±0.8 ±0.4
100
μL
±0.8 ±0.8 ±0.3 ±0.3
30 μL – 300 μL
Orange
20 μL – 300 μL
55 mm
30
μL
±3.0 ±0.9 ±1.0 ±0.3
150
μL
±1.0 ±1.5 ±0.5 ±0.75
300
μL
±0.6 ±1.8 ±0.3 ±0.9