Adjustment
Eppendorf SOP
English (EN)
7.2.4
Reference 2 adjustable-volume
7.2.4.1 User adjustment
Changing the adjustment will change the volume by a specific value. Strictly
speaking, the change only applies to the testing volume.
Example:
You readjust a 10 – 100 μL pipette, with a volume setting of 100 μL, by 1 μL
(1 μL ≙ 1%). If the volume setting is 10 μL, the pipette is also adjusted by 1 μL
(≙ 10%).
Auxiliary aids from the delivery package
• Pin
• Red plastic adjustment seal (ADJ)
1. Insert the pin in the middle of the gray adjustment seal (ADJ).
2. Remove the adjustment seal.
3. Insert the adjustment tool.
4. Turn the adjustment tool until the adjustment display shows the desired
value.
6. Carry out weighings to check the accuracy and precision.
5. Read the set value distortion-free using the aligning aid in the viewing
window.