4 research plus fixed, Research plus fixed – Eppendorf SOP User Manual
Page 40

40
Technical specifications
Eppendorf SOP
English (EN)
9.1.4
Research plus fixed
Research plus fixed volume
Model
Test tip
epT.I.P.S. Color
code
Volume range
Length
Eppendorf AG error limits
Error
Systematic
error
Random error
± %
± µL
± %
± µL
10 μL
Medium gray
0.1 μL – 20 μL
40 mm
±1.2
±0.12 ±0.6 ±0.06
20 μL
light gray
0.5 μL – 20 μL L
46 mm
±0.8
±0.16
±0.3
±0.06
10 μL
Yellow
2 μL - 200 μL
53 mm
±1.2
±0.12
±0.6
±0.06
20 μL
±1.0
±0.2
±0.3
±0.06
25 μL
±1.0
±0.25
±0.3
±0.08
50 μL
±0.7
±0.35
±0.3
±0.15
100 μL
±0.6
±0.6
±0.2
±0.2
200 μL
±0.6
±1.2
±0.2
±0.4
200 μL
Blue
50 μL – 1 000 μL
71 mm
±0.6
±1.2
±0.2
±0.4
250 μL
±0.6
±1.5
±0.2
±0.5
500 μL
±0.6
±3.0
±0.2
±1.0
1,000 μL
±0.6
±6.0
±0.2
±2.0