3 research fixed, Research fixed – Eppendorf SOP User Manual
Page 39

39
Technical specifications
Eppendorf SOP
English (EN)
9.1.3
Research fixed
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Error limits
Error
Systematic error
Random error
± %
± µL
± %
± µL
10 μL
Yellow
2 - 200 μL
53 mm
±1.2
±0.12
±0.6
±0.06
20 μL
±1.0
±0.2
±0.3
±0.06
25 μL
±1.0
±0.25
±0.3
±0.075
50 μL
±0.7
±0.35
±0.3
±0.15
100 μL
±0.6
±0.6
±0.2
±0.2
200 μL
Blue
0.05 - 1 mL
71 mm
±0.6
±1.2
±0.2
±0.4
250 μL
±0.6
±1.5
±0.2
±0.5
500 μL
±0.6
±3.0
±0.2
±1.0
1,000 μL
±0.6
±6.0
±0.2
±2.0