3 research plus, Research plus – Eppendorf SOP User Manual
Page 51
51
Technical specifications
Eppendorf SOP
English (EN)
9.3.3
Research plus
Research plus adjustable multi-channel
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Eppendorf AG error limits
Error
Systematic
error
Random
error
± %
± µL
± %
± µL
0.5 μL – 10 μL
Increment:
0.01 μL
Medium gray
0.1 μL – 20 μL
40 mm
0.5 μL ±12
±0.06 ±8.0
±0.04
1 μL
±8.0 ±0.08 ±5.0
±0.05
5 μL
±4.0 ±0.2
±2.0
±0.1
10 μL
±2.0 ±0.2
±1.0
±0.1
10 μL – 100 μL
Increment:
0.1 μL
Yellow
2 μL - 200 μL
53 mm
10 μL
±3.0 ±0.3
±2.0
±0.2
50 μL
±1.0 ±0.5
±0.8
±0.4
100 μL ±0.8 ±0.8
±0.3
±0.3
30 μL – 300 μL
Increment:
0.2 μL
Orange
20 μL – 300 μL
55 mm
30 μL
±3.0 ±0.9
±1.0
±0.3
150 μL ±1.0 ±1.5
±0.5
±0.75
300 μL ±0.6 ±1.8
±0.3
±0.9