3 multi-channel pipettes, 1 research, Multi-channel pipettes 9.3.1 – Eppendorf SOP User Manual
Page 49: Research

49
Technical specifications
Eppendorf SOP
English (EN)
9.3
Multi-channel pipettes
9.3.1
Research
Model
Test tip
epT.I.P.S. Color
code
Volume range
Length
Volume
in µL
Error limits
Error
Systematic
error
Random error
± %
± µL
± %
± µL
Research8-channel
0.5 - 10 μL
light gray
0.5 – 20 μL
46 mm
1
±8.0 ±0.08
±5.0 ±0.05
5
±4.0 ±0.2 ±2.0 ±0.1
10
±2.0 ±0.2 ±1.0 ±0.1
Research12-channel
0.5 - 10 μL
see 8-channel
Research8-channel
10 - 100 μL
Yellow
2 - 200 μL
53 mm
10
±3.0 ±0.3 ±2.0 ±0.2
50
±1.0 ±0.5 ±0.8 ±0.4
100 ±0.8 ±0.8 ±0.3 ±0.3
Research12-channel
10 - 100 μL
see 8-channel
Research8-channel
30 - 300 μL
Orange
20 – 300 μL
55 mm
30
±3.0 ±0.9 ±1.0 ±0.3
150 ±1.0 ±1.5 ±0.5 ±0.75
300 ±0.6 ±1.8 ±0.3 ±0.9
Research12-channel
30 - 300 μL
see 8-channel