2 multipettem4, Multipettem4 – Eppendorf SOP User Manual
Page 55

55
Technical specifications
Eppendorf SOP
English (EN)
9.4.2
MultipetteM4
Combitip advanced
Testing volume
Error limits
Error
Systematic error Random error
± %
± µL
± %
± µL
0.1 mL
white
Increment: 1 μL
2 μL
±1.6
±0.032
±3.0
±0.06
20 μL
±1.0
±0.2
±2.0
±0.4
0.2 mL
light blue
Increment: 2 μL
4 μL
±1.3
±0.052
±2.0
±0.08
40 μL
±0.8
±0.32
±1.5
±0.6
0.5 mL
purple
Increment: 5 μL
10 μL
±0.9
±0.09
±1.5
±0.15
100 μL
±0.8
±0.8
±0.6
±0.6
1 mL
yellow
Increment: 10 μL
20 μL
±0.9
±0.18
±0.9
±0.18
200 μL
±0.6
±1.2
±0.4
±0.8
2.5 mL
green
Increment: 25 μL
50 μL
±0.8
±0.4
±0.8
±0.4
500 μL
±0.5
±2.5
±0.3
±1.5
5 mL
blue
Increment: 50 μL
100 μL
±0.6
±0.6
±0.6
±0.6
1 000 μL
±0.5
±5.0
±0.25
±2.5
10 mL
orange
Increment: 0.1 mL
200 μL
0.2 mL
±0.5 ±1.0
±0.6
±1.2
2 000 μL
2 mL
±0.5 ±10
±0.25
±5.0
25 mL
red
Increment: 0.25 mL
500 μL
0.5 mL
±0.4 ±2.0
±0.6
±3.0
5 000 μL
5 mL
±0.3 ±15
±0.25
±12.5
50 mL
light gray
Increment: 0.5 mL
1 000 μL
1 mL
±0.3 ±3.0
±0.5
±5.0
10 000 μL
10 mL
±0.3 ±30
±0.3
±30