5 xplorer and xplorer plus, Xplorer and xplorer plus – Eppendorf SOP User Manual
Page 53

53
Technical specifications
Eppendorf SOP
English (EN)
9.3.5
Xplorer and Xplorer plus
pipette
Increment
Pipette tip
Color code
Volume range
Length
Testing
volume
Error limits
Error
Systematic
error
Random error
µL
µL
%
µL
%
µL
0.5 μL – 10 μL
Increment:
0.01 μL
Medium gray
0.1 μL – 20 μL
40 mm
1 μL
±5.0
±0.05
±3.0
±0.03
5 μL
±3.0
±0.15
±1.5
±0.075
10 μL
±2.0
±0.2
±0.8
±0.08
5 μL – 100 μL
Increment:
0.1 μL
Yellow
2 μL –200 μL
53 mm
10 μL
±2.0
±0.2
±2.0
±0.2
50 μL
±1.0
±0.5
±0.8
±0.4
100 μL
±0.8
±0.8
±0.25
±0.25
15 μL – 300 μL
Increment:
0.2 μL
Orange
15 μL –300 μL
55 mm
30 μL
±2.5
±0.75
±1.0
±0.3
150 μL
±1.0
±1.5
±0.5
±0.75
300 μL
±0.6
±1.8
±0.25
±0.75
50 μL – 1200 μL
Increment:
1 μL
Green
50 μL–1250 μL
76 mm
120 μL
±6.0
±7.2
±0.9
±1.08
600 μL
±2.7
±16.2
±0.4
±2.4
1200 μL ±1.2
±14.2
±0.3
±3.6