Technical specifications
Eppendorf SOP
English (EN)

9.4

Multipette

The following specifications for the Multipette plus, Multipette stream /
Multipette Xstream, Multipette M4 apply to the following conditions:
• Using the Combitips advanced
• Liquid: Distilled or deionized water
• Reference temperature: 20 to 25 °C, ±0.5 °C
• Number of determinations: 10 according to EN ISO 8655, with the original

Eppendorf Combitip advanced

Multipette stream / Multipette Xstream:
• Volume test in "DIS" mode
• Set speed level: 7

9.4.1

MultipettePlus

Combitip advanced Testing volume

Error limits

Error

Systematic error

Random error

± %

± µL

± %

± µL

0.1 mL
(beige piston)

2 μL

±1.6

±0.032 ±3.0

±0.06

20

μL

±1.0 ±0.2 ±2.0 ±0.4

0.2 mL
(blue piston)

4 μL

±1.3

±0.052 ±2.0

±0.08

40 μL

±0.8

±0.32

±1.5

±0.6

0.5 mL

10 μL

±0.9

±0.09

±1.5

±0.15

100

μL

±0.8 ±0.8 ±0.6 ±0.6

1 mL

20 μL

±0.9

±0.18

±0.9

±0.18

200

μL

±0.6 ±1.2 ±0.4 ±0.8

2.5 mL

50 μL

±0.8

±0.4

±0.8

±0.4

500

μL

±0.5 ±2.5 ±0.3 ±1.5

5

mL

100

μL

±0.6 ±0.6 ±0.6 ±0.6

1,000

μL

±0.5 ±5.0 ±0.25 ±2.5

10

mL

200

μL

±0.5 ±1.0 ±0.6 ±1.2

2 000 μL

±0.5

±10

±0.25

±5.0

25 mL
(blue adapter)

500

μL

±0.4 ±2.0 ±0.6 ±3.0

5,000 μL

±0.3

±15

±0.25

±12.5

50 mL
(dark gray adapter)

1,000

μL

±0.3 ±3.0 ±0.5 ±5.0

10,000 μL

±0.3

±30

±0.3

±30