2 reference2, 3 research fixed, Reference2 – Eppendorf SOP User Manual
Page 34: Research fixed

34
Adjustment
Eppendorf SOP
English (EN)
7.3.2
Reference2
Adjustment is carried out as described for the Reference 2 adjustable-volume
(see p. 20).
7.3.3
Research fixed
Auxiliary equipment
• Supplied pipette wrench (order no. 3111 501.016)
With reference to water, one revolution of the volume setting ring corresponds to:
1. Insert side D of the pipette wrench
vertically into the lateral adjustment
opening of the pipette grip.
2. Move the tool into the vertical
position.
3. Turn the volume setting ring toward
– or +. This adjusts the piston stoke
of the pipette.
Volume range
Vol./turn
10
approx. 0.8 μL
20
approx. 0.8 μL
25
approx. 0.8 μL
50
approx. 0.8 μL
100
approx. 0.8 μL
200
approx. 38 μL
250
approx. 38 μL
500
approx. 38 μL
1 000
approx. 38 μL