Campbell Scientific CR23X Micrologger User Manual
Page 206

SECTION 13. CR23X MEASUREMENTS
13-6
DETERMINING SOURCE RESISTANCE
The source resistance used to estimate the
settling time constant is the resistance the
CR23X input "sees" looking out at the sensor.
For our purposes the source resistance can be
defined as the resistance from the CR23X input
through all external paths back to the CR23X.
Figure 13.3-2 shows a typical resistive sensor,
(e.g., a thermistor) configured as a half bridge.
Figure 13.3-3 shows Figure 13.3-2 re-drawn in
terms of the resistive paths determining the
source resistance Ro, is given by the parallel
resistance of Rs and Rf, as shown in Equation
13.3-8.
R
o
= R
s
R
f
/(R
s
+R
f
)
[13.3-8]
If R
f
is much smaller, equal to or much greater
than R
s
, the source resistance can be
approximated by Equations 13.3-9 through
13.3-11, respectively.
R
o
~
R
f
, R
f
< s [13.3-9] R o = R f /2, R f =R s [13.3-10] R o ~ R s , R f >>R s [13.3-11] The source resistance for several Campbell DETERMINING LEAD CAPACITANCE Wire manufacturers typically provide two f ) in single-ended measurements such as Figure 13.3-2, the w , usually specified as pfd/ft. C w is actually the sum of capacitance between the two conductors and the FIGURE 13.3-4. Wire Manufacturers Capacitance Specifications, C w TABLE 13.3-2. Properties of Lead Wires Used by Campbell Scientific CSI Rl C w Part # Conductors Insulation AWG (ohms/1000ft.) (pfd/ft.) 9721 3 shld. 3 pair polypropylene 24 28 41 9661 1 shld. pair polypropylene 22 18 42 7359 1 shld. 4 cond. polyethylene 22 16 32
Scientific sensors are given in column 3 of
Table 13.3-5.
capacitance specifications: 1) the capacitance
between the two leads with the shield floating,
and 2) the capacitance between the two leads
with the shield tied to one lead. Since the input
lead and the shield are tied to ground (often
through a bridge resistor, R
second specification is used in determining lead
capacitance. Figure 13.3-4 is a representation
of this capacitance, C
capacitance between the top conductor and the
shield. Capacitance for 3 lead wires used in
Campbell Scientific sensors is shown in column
6 of Table 13.3-2.