SECTION 8. PROCESSING AND PROGRAM CONTROL EXAMPLES

8-12

Table 8.10-3 lists the input channel
configuration and Input Storage allocation for
the measured values. After reading the new
input samples, the Level 2 measurements are
relocated using the Block Move Instruction 54,
then Ta1 is relocated through a separate move
and e1 is positioned by specifying the
destination location in the Wet/Dry-Bulb
Instruction. The COV/CORR Instruction must
be entered twice, once for each level.

In addition to ordering Level 1 and Level 2 in
locations 1-5 and 11-15 respectively, 2 more
locations are required. Converting the wet-/dry-

bulb measurements to vapor pressure using
Instruction 57 requires atmospheric pressure.
We'll use the standard atmosphere for the site
elevation and key the value into Location 17
using the C command in the

6

Mode.

This example requires that 54 locations be
allotted to Input Storage and 79 to Intermediate
Storage (35 for the 1st COV/CORR Instruction,
43 for the second, and 1 for Instruction 92).

TABLE 8.10-3. Example Input Channel and Location Assignments

INPUT

INPUT

INPUT

INPUT

PARAM

CHAN

LOC

PARAM LOC

PARA LOC

W1

1

1

W1

1

W1

1

U1

2

2

U1

2

U1

2

V1

3

3

V1

3

V1

3

W2

4

4

Ta1

9 ---------------------- Ta1

4

U2

5

5

Tw1

10

Separate moves e1

5

V2

6

6

W2

11

W2

11

Ta2

7

7

->Block->

U2

12

U2

12

Tw2

8

8

move

V2

13

V2

13

Ta1

9

9

Ta2

14

Ta2

14

Tw1

10

10

Tw2

15

e2

15

PROGRAM

*

Table 1 Program

01:

1

Execution Interval (seconds)

01:

Set Port(s) (P20)
1:

9999

C8..C5 = nc/nc/nc/nc

2:

9933

C4..C1 = nc/nc/1ms/1ms

02:

Volts (SE) (P1)
1:

6

Reps

2:

5

±5000 mV Slow Range

3:

1

SE Channel

4:

1

Loc [ W1 ]

5:

.018

Mult

6:

0

Offset