4 installation instructions, 1 incoming acceptance, transport, storage, 1 incoming acceptance – Pyromation T82 User Manual
Page 8: 2 transport and storage, 2 installation conditions, 1 dimensions, 2 installation point, 3 installation instructions, Ä 8), Installation instructions
Installation instructions
T82
8
4
Installation instructions
4.1
Incoming acceptance, transport, storage
4.1.1
Incoming acceptance
• Is the packaging or content damaged?
• Is the delivery complete and is anything missing? Check the scope of delivery against your
order.
4.1.2
Transport and storage
• Pack the device in such a way as to protect it reliably against impact for storage (and
transportation). The original packaging provides optimum protection.
• The permitted storage temperature is -40 to +100 °C (-40 to 212 °F).
4.2
Installation conditions
4.2.1
Dimensions
The dimensions of the device are provided in the 'Technical data' section (® ä 33).
4.2.2
Installation point
• In the terminal head, flat face, as per DIN 43729, direct mounting on insert with cable entry
(middle hole 7 mm)
• In the field housing, separated from the process
It is also possible to mount the device on a top-hat rail as per EN 60715 using the DIN rail
clip accessory .
Information about the conditions (such as the ambient temperature, degree of protection, climate
class etc.) that must be present at the measuring point so that the device can be mounted
correctly is provided in the 'Technical data' section (® ä 33).
When using in the hazardous area, the limit values of the certificates and approvals must be
observed (see Ex-Safety Instructions).
4.3
Installation instructions
A Phillips head screwdriver is required to mount the head transmitter.
NOTICE
Do not overtighten the mounting screws as this could damage the head transmitter.
►
Maximum torque = 1 Nm (¾ pound-feet).