3 "output" submenu, 4 "communication" submenu, Ä 77) – Pyromation T82 User Manual
Page 77
![background image](/manuals/370052/77/background.png)
T82
Operating menu and parameter description
77
13.4.3 "Output" submenu
Measuring mode
Navigation
Expert ® Output ® Measuring mode
Description
Enables the inversion of the output signal.
Additional information
•
Standard
The output current increases with increasing temperatures
•
Inverse
The output current decreases with increasing temperatures
Options:
• Standard
• Inverse
Factory settings
Standard
13.4.4 "Communication" submenu
"HART configuration" submenu
Navigation
Expert ® Communication ® HART configuration ® Device tag
HART short tag
Navigation
Expert ® Communication ® HART configuration ® HART short tag
Description
Use this function to define a short tag for the measuring point.
User input
Up to 8 alphanumeric characters (letters, numbers and special characters)
Factory settings
SHORTTAG
HART address
Navigation
Expert ® Communication ® HART configuration ® HART address
Description
Use this function to define the HART address of the device.