2 submenus and user roles, T82 operating options 19 – Pyromation T82 User Manual
Page 19
T82
Operating options
19
6.2.2
Submenus and user roles
Certain parts of the menu are assigned to certain user roles. Each user role corresponds to typical
tasks within the lifecycle of the device.
User role
Typical tasks
Menu
Content/meaning
Operator
Tasks during operation:
• Configuration of the display.
• Reading measured values.
"Display/operation" Contains all the parameters that are required in ongoing
operation: configuration of the measured value display
(displayed values, display format, etc.).
Maintenance
Commissioning:
• Configuration of the measurement.
• Configuration of data processing (scaling, linearization,
etc.).
• Configuration of the analog measured value output.
"Setup"
Contains all parameters for commissioning:.
•
Setup parameters
Once values have been set for these parameters, the
measurement should generally be completely configured.
•
"Advanced setup" submenu
Contains additional submenus and parameters:
– For more accurate configuration of the measurement
(adaptation to special measuring conditions).
– For converting the measured value (scaling,
linearization).
– For scaling the output signal.
Fault elimination:
• Diagnosing and eliminating process errors.
• Interpretation of device error messages and correcting
associated errors.
"Diagnostics"
Contains all parameters for detecting and analyzing errors:
•
Diagnostic list
Contains up to 3 currently pending error messages.
•
Event logbook
Contains the 5 most recent error messages (no longer
pending).
•
"Device information" submenu
Contains information for identifying the device.
•
"Measured values" submenu
Contains all current measured values.
•
"Simulation" submenu
Is used to simulate measured values or output values.
•
"Device reset" submenu
Expert
Tasks that require detailed knowledge of the function of
the device:
• Commissioning measurements under difficult
conditions.
• Optimal adaptation of the measurement to difficult
conditions.
• Detailed configuration of the communication interface.
• Error diagnostics in difficult cases.
"Expert"
Contains all parameters of the device (including those that
are already in one of the other menus). This menu is
structured according to the function blocks of the device:
•
"System" submenu
Contains all higher-order device parameters that do not
pertain either to measurement or the measured value
communication.
•
"Sensors" submenu
Contains all parameters for configuring the
measurement.
•
"Output" submenu
Contains all parameters for configuring the analog
current output.
•
"Communication" submenu
Contains all parameters for configuring the digital
communication interface.
•
"Diagnostics" submenu
Contains all parameters for detecting and analyzing
errors.