"Sensor" submenu
Sensor offset n
n = Stands for the number of sensor inputs (1 and 2)
Navigation
Setup ® Advanced setup ® Sensor ® Sensor offset n
Expert ® Sensor ® Sensor n ® Sensor offset n
Description
Use this function to set the zero point correction (offset) of the sensor measured value. The value
indicated is added to the measured value.
User input
-10.0...+10.0
Factory settings
0.0
Corrosion detection
Navigation
Setup ® Advanced setup ® Sensor ® Corrosion detection
Expert ® Sensor ® Diagnostic settings ® Corrosion detection
Description
Use this function to select the category (status signal) which is displayed when corrosion of the
sensor connection cables is detected.
Only possible for RTD sensors with 4-wire connection and thermocouples (TC).
Options:
• Maintenance required (M)
• Failure (F)
Factory settings
Maintenance required (M)
Drift/difference mode
Navigation
Setup ® Advanced setup ® Sensor ® Drift/difference mode
Expert ® Sensor ® Diagnostic settings ® Drift/difference mode
Description
Use this function to choose whether the device reacts to the value exceeding or dropping below
the drift/difference set point.
Can only be selected for 2-channel operation.
Additional information
• If the Out band (drift) option is selected, a status signal is displayed if the absolute value for
the differential value exceeds the drift/difference set point.
• If the In band option is selected, a status signal is displayed if the absolute value for the
differential value drops below the drift/difference set point.