beautypg.com

Ä 57) – Pyromation T82 User Manual

Page 57

background image

T82

Operating menu and parameter description

57

"Sensor" submenu

Sensor offset n

n = Stands for the number of sensor inputs (1 and 2)

Navigation

Setup ® Advanced setup ® Sensor ® Sensor offset n
Expert ® Sensor ® Sensor n ® Sensor offset n

Description

Use this function to set the zero point correction (offset) of the sensor measured value. The value
indicated is added to the measured value.

User input

-10.0...+10.0

Factory settings

0.0

Corrosion detection

Navigation

Setup ® Advanced setup ® Sensor ® Corrosion detection
Expert ® Sensor ® Diagnostic settings ® Corrosion detection

Description

Use this function to select the category (status signal) which is displayed when corrosion of the
sensor connection cables is detected.

Only possible for RTD sensors with 4-wire connection and thermocouples (TC).

Options:

• Maintenance required (M)
• Failure (F)

Factory settings

Maintenance required (M)

Drift/difference mode

Navigation

Setup ® Advanced setup ® Sensor ® Drift/difference mode
Expert ® Sensor ® Diagnostic settings ® Drift/difference mode

Description

Use this function to choose whether the device reacts to the value exceeding or dropping below
the drift/difference set point.

Can only be selected for 2-channel operation.

Additional information

• If the Out band (drift) option is selected, a status signal is displayed if the absolute value for

the differential value exceeds the drift/difference set point.

• If the In band option is selected, a status signal is displayed if the absolute value for the

differential value drops below the drift/difference set point.