Ä 74) – Pyromation T82 User Manual
Page 74
Operating menu and parameter description
T82
74
8. If special linearization is also used for a second sensor, repeat steps 2 to 6.
ß
9. End
Sensor n lower limit
Navigation
Expert ® Sensor® Sensor n ® Linearization ® Sensor n lower limit
Prerequisite
The RTD platinum, RTD poly nickel or RTD polynomial copper option is enabled in the Sensor
type parameter.
Description
Use this function to set the lower calculation limit for special sensor linearization.
User input
Depends on the sensor type selected.
Factory settings
-200 °C
Sensor n upper limit
Navigation
Expert ® Sensor ® Sensor n ® Linearization ® Sensor n upper limit
Prerequisite
The RTD platinum, RTD poly nickel or RTD polynomial copper option is enabled in the Sensor
type parameter.
Description
Use this function to set the upper calculation limit for special sensor linearization.
User input
Depends on the sensor type selected.
Factory settings
850 °C
Call./v. Dusen coeff. R0
Navigation
Expert ® Sensor ® Sensor n ® Linearization ® Call./v. Dusen coeff. R0
Prerequisite
The RTD platinum (Callendar/Van Dusen) option is enabled in the Sensor type parameter.
Description
Use this function to set the R0 value for linearization with the Callendar/Van Dusen polynomial.
User input
40.000 to 1
050.000
Factory settings
100.000 Ohm