beautypg.com

Ä 74) – Pyromation T82 User Manual

Page 74

background image

Operating menu and parameter description

T82

74

8. If special linearization is also used for a second sensor, repeat steps 2 to 6.

ß

9. End

Sensor n lower limit

Navigation

Expert ® Sensor® Sensor n ® Linearization ® Sensor n lower limit

Prerequisite

The RTD platinum, RTD poly nickel or RTD polynomial copper option is enabled in the Sensor
type parameter.

Description

Use this function to set the lower calculation limit for special sensor linearization.

User input

Depends on the sensor type selected.

Factory settings

-200 °C

Sensor n upper limit

Navigation

Expert ® Sensor ® Sensor n ® Linearization ® Sensor n upper limit

Prerequisite

The RTD platinum, RTD poly nickel or RTD polynomial copper option is enabled in the Sensor
type parameter.

Description

Use this function to set the upper calculation limit for special sensor linearization.

User input

Depends on the sensor type selected.

Factory settings

850 °C

Call./v. Dusen coeff. R0

Navigation

Expert ® Sensor ® Sensor n ® Linearization ® Call./v. Dusen coeff. R0

Prerequisite

The RTD platinum (Callendar/Van Dusen) option is enabled in the Sensor type parameter.

Description

Use this function to set the R0 value for linearization with the Callendar/Van Dusen polynomial.

User input

40.000 to 1

050.000

Factory settings

100.000 Ohm