6 "device reset" submenu, Ä 69), Д 69) – Pyromation T82 User Manual
Page 69
T82
Operating menu and parameter description
69
Additional information
The simulation value is defined in the Value current output parameter.
Value current output
Navigation
Diagnostics ® Simulation ® Value current output
Expert ® Diagnostics ® Simulation ® Value current output
Additional information
The Simulation current output parameter must be set to On.
Description
Use this function to set a current value for the simulation. In this way, users can verify the correct
adjustment of the current output and the correct function of downstream switching units.
User input
3.59 to 23.0 mA
Factory settings
3.59 mA
13.3.6 "Device reset" submenu
Device reset
Navigation
Diagnostics ® Device reset ® Device reset
Expert ® Diagnostics ® Device reset ® Device reset
Description
Use this function to reset the device configuration - either entirely or in part - to a defined state.
Options:
•
Not active
No action is executed and the user exits the parameter.
•
To factory defaults
All the parameters are reset to the factory setting.
•
To delivery settings
All the parameters are reset to the order configuration.The order configuration can differ from
the factory setting if customer-specific parameter values were defined when the device was
ordered.
•
Restart device
The device is restarted but the device configuration remains unchanged.
Factory settings
Not active