T82 diagnostics and troubleshooting 31 – Pyromation T82 User Manual
Page 31
T82
Diagnostics and troubleshooting
31
Diagno
stics
number
Event text
Remedial measures
Status
signal
from the
factory
Event
level from
the
factory
Changeab
le in
102
Sensor value too
high
1. Check process temperature.
2. Check sensor.
3. Check sensor type.
F
Alarm
104
Backup active
1. Check electrical connection sensor 1.
2. Replace sensor 1.
3. Check connection type.
M
Warning
105
Calibration interval 1. Execute calibration and reset calibration interval.
2. Switch off calibration counter.
M
Warning
F
106
Backup not
available
1. Check electrical connection sensor 2.
2. Replace sensor 2.
3. Check connection type.
M
Warning
Diagnostics for the electronics
201
Electronic error
Replace electronics.
F
Alarm
221
Electronic
reference
Replace electronics.
F
Alarm
241
Electronic software 1. Device restart.
2. Device reset.
3. Replace electronics.
F
Alarm
261
Electronic modules Replace electronics.
F
Alarm
262
Module connection 1. Check whether the retainers and the connection of the
display module are correctly seated on the head transmitter.
2. Test the display module with other suitable head
transmitters.
3. Display module defective? Replace module.
M
Warning
283
Memory content
Replace electronics.
F
Alarm
301
Supply voltage
1. Increase supply voltage.
2. Check electrical connection for corrosion.
F
Alarm
Diagnostics for the configuration
401
Factory reset
Please wait during the reset process.
C
Warning
402
Configuration
initialization
Please wait during the initialization process
C
Warning
411
Up-/Download
Please wait during the Up-/Download process.
C
Warning
431
Factory calibration
Replace electronics.
F
Alarm
435
Linearization
1. Check configuration of sensor parameters.
2. Check configuration of special sensor linearization.
3. Contact service organisation.
4. Replace electronics.
F
Alarm
437
Configuration
1. Check configuration of sensor parameters.
2. Check configuration of special sensor linearization.
3. Check transmitter settings.
4. Contact service organisation.
F
Alarm
451
Data handling
Please wait during the data handling process.
C
Warning
483
Simulation input
Deactivate simulation.
C
Warning
485
Simulation
measured value
491
Simulation current
output