beautypg.com

4 "expert" menu, 1 "system" submenu, Ä 70) – Pyromation T82 User Manual

Page 70

background image

Operating menu and parameter description

T82

70

13.4

"Expert" menu

The parameter groups for the Expert setup contain all the parameters of the "Display/operation",
"Setup" and "Diagnostics" operating menus, as well as other parameters that are solely reserved
for experts.

13.4.1 "System" submenu

Damping

Navigation

Expert ® System ® Damping

Description

Use this function to set the time constant for current output damping.

User input

0 to 120 s

Factory settings

0.00 s

Additional information

The current output reacts with an exponential delay to fluctuations in the measured value. The
time constant of this delay is specified by this parameter. If a low time constant is entered, the
current output reacts quickly to the measured value. On the other hand, if a high time constant is
entered, the current output reaction is delayed.

Alarm delay

Navigation

Expert ® System ® Alarm delay

Description

Use this function to set the delay time during which a diagnostics signal is suppressed before it is
output.

User input

0 to 5 s

Factory settings

2 s

Mains filter

Navigation

Expert ® System ® Mains filter

Description

Use this function to select the mains filter for A/D conversion.

Options:

• 50Hz
• 60Hz

Factory setting

50Hz