beautypg.com

Ä 67) – Pyromation T82 User Manual

Page 67

background image

T82

Operating menu and parameter description

67

"Min/max values" submenu

Sensor n min value

n = Stands for the number of sensor inputs (1 and 2)

Navigation

Diagnostics ® Measured values ® Min/max values ® Sensor n min value
Expert ® Diagnostics ® Measured values ® Min/max values ® Sensor n min value

Description

Use this function to display the minimum temperature measured in the past at sensor input 1 or 2
(peakhold indicator).

Sensor n max value

n = Stands for the number of sensor inputs (1 and 2)

Navigation

Diagnostics ® Measured values ® Min/max values ® Sensor n max value
Expert ® Diagnostics ® Measured values ® Min/max values ® Sensor n max value

Description

Use this function to display the maximum temperature measured in the past at sensor input 1 or
2 (maximum indicator).

Reset sensor min/max values

Navigation

Diagnostics ® Measured values ® Min/max values ® Reset sensor min/max values
Expert ® Diagnostics ® Measured values ® Min/max values ® Reset sensor min/max
values

Description

Reset the maximum indicators for the minimum and maximum temperatures measured at the
sensor inputs.

Options:

• No
• Yes

Factory setting

No

Device temperature max.

Navigation

Diagnostics ® Measured values ® Min/max values ® Device temperature max.
Expert ® Diagnostics ® Measured values ® Min/max values ® Device temperature
max.