Ä 59) – Pyromation T82 User Manual
Page 59
T82
Operating menu and parameter description
59
Additional information
The threshold value is relevant if the sensor switching function is assigned to a HART
®
variable
(PV, SV, TV, QV).
Options:
Depends on the sensor types selected.
Factory settings
0.0 °C
"Current output" submenu
Adjustment of the analog output (4 and 20 mA current trimming)
Current trimming is used to compensate the analog output (D/A conversion). Here, the output
current of the transmitter can be adapted so that it suits the value expected at the higher-order
system.
NOTICE
Current trimming does not affect the digital HART
®
value. This can cause the measured
value shown on the plug-in display to differ from the value displayed in the higher-
order system.
►
The digital measured values can be adapted with the sensor trimming parameter in the menu
Expert ® Sensor ® Sensor trimming.
Procedure
1. Start
ß
2. Install an accurate amperemeter (more accurate than the transmitter) in the current loop.
ß
3. Switch on current output simulation and set the simulation value to 4 mA.
ß
4. Measure the loop current with the amperemeter and make a note of the value.
ß
5. Set the simulation value to 20 mA.
ß
6. Measure the loop current with the amperemeter and make a note of the value.
ß
7. Enter the current values determined as adjustment values in the Current trimming 4 mA / 20 mA parameters.
ß
8. End
Output current
Navigation
Setup ® Advanced setup ® Current output® Output current
Expert ® Output ® Output current
Description
Use this function to view the calculated output current in mA.