Technical data
T82
38
Altitude
Up to 4000 m (4374.5 yards) above mean sea level as per IEC 61010-1, CAN/CSA C22.2 No.
61010-1
Climate class
As per IEC 60654-1, Class C
Humidity
• Condensation permitted as per IEC 60 068-2-33
• Max. rel. humidity: 95% as per IEC 60068-2-30
Degree of protection
IP 20. In the installed state, depends on the terminal head or field housing used.
Vibration
25 to 100 Hz for 4g (increased vibration stress) as per GL-guidelines, chapter 2, edition 2003
Electromagnetic compatibility
(EMC)
CE compliance
Electromagnetic compatibility in accordance with all the relevant requirements of the EN 61326
series and NAMUR Recommendation EMC (NE21). Details are provided in the Declaration of
Conformity. All tests were passed both with and without ongoing digital HART
®
communication.
ESD (electrostatic discharge)
EN/IEC 61000-4-2
6 kV cont., 8 kV air
Electromagnetic fields
EN/IEC 61000-4-3
0.08 to 2.7 GHz
10 V/m
Burst (fast transients)
EN/IEC 61000-4-4
2 kV
Surge (surge voltage)
EN/IEC 61000-4-5
0.5 kV sym.
1 kV assym.
Conducted RF
EN/IEC 61000-4-6
0.01 to 80 MHz
10 V
Measuring category
Measuring category II as per IEC 61010-1. The measuring category is provided for measuring on
power circuits that are directly connected electrically with the low-voltage network.
Degree of contamination
Pollution degree 2 as per IEC 61010-1.
12.6
Mechanical construction
Design, dimensions
24.1 (0.95
”)
33 (1.3
”)
Ш
44 (1.73
”)
Ш
7 (0.28
”)
Ш5 (0.2”)
B
C
A
A0007301
å
11
Version with screw terminals, dimensions in mm (in).
A
Spring travel L ≥ 5 mm (not for US - M4 securing screws)
B
Fasteners for attachable measured value display
C
Interface for contacting the measured value display