beautypg.com

1 "display/operation" menu, Ä 47) – Pyromation T82 User Manual

Page 47

background image

T82

Operating menu and parameter description

47

13.1

"Display/operation" menu

The settings for displaying the measured value on the optional plug-in display are made in the
"Display/Operation" menu. The following parameters can be found in "Display/operation" and
"Expert ® System ® Display".

These settings do not have any effect on the output values of the transmitter. They are only
used to configure how information is shown on the display.

Display interval

Navigation

Display/operation ® Display interval
Expert ® System ® Display ® Display interval

Description

Use this function to set the length of time the measured values are displayed if the values
alternate on the display. The display only alternates between values if more than one measured
value is defined.

• The Value 1 display - Value 3 display parameters are used to specify what measured

values are shown on the display (® ä 48).

• The display format of the displayed measured values is specified using the Format display

parameter.

User input

4 to 20 s

Factory settings

4 s

Format display

Navigation

Display/operation ® Format display
Expert ® System ® Display ® Format display

Description

Use this function to select how the measured value is shown on the local display. The display
format Measured value or Measured value with bar graph can be configured.

Options:

• Value only
• Value + Bargraph

Factory settings

Value only