Ä 76) – Pyromation T82 User Manual
Page 76

Operating menu and parameter description
T82
76
Description
Option to control the calibration counter.
• The countdown duration (in days) is specified with the Calibration counter start value
parameter.
• The status signal issued when the limit value is reached is defined with the Calibration
counter alarm category parameter.
Options:
• Off: Stops the calibration counter
• On: Starts the calibration counter
• Reset + run: Resets to the set start value and starts the calibration counter
Factory settings
Off
Calibration counter alarm category
Navigation
Expert ® Sensor ® Diagnostic settings ® Calibration counter alarm category
Description
Use this function to select the category (status signal) as to how the device reacts to the set
calibration countdown.
Options:
• Maintenance required (M)
• Failure (F)
Factory settings
Maintenance required (M)
Calibration counter start value
Navigation
Expert ® Sensor ® Diagnostic settings ® Calibration counter start value
Description
Use this function to set the start value for the calibration counter.
User input
0 to 365 d (days)
Factory settings
365
Calibration countdown
Navigation
Expert ® Sensor ® Diagnostic settings ® Calibration countdown
Description
Use this function to view the time remaining until the next calibration.
The countdown of the calibration counter is only running if the device is switched on.
Example: The calibration counter is set to 365 days on January 1st, 2011. If the device will
be switched off for 100 days, the calibration counter alarm is displayed on April 10th, 2012.