Operating menu and parameter description

T82

76

Description

Option to control the calibration counter.

• The countdown duration (in days) is specified with the Calibration counter start value

parameter.

• The status signal issued when the limit value is reached is defined with the Calibration

counter alarm category parameter.

Options:

• Off: Stops the calibration counter
• On: Starts the calibration counter
• Reset + run: Resets to the set start value and starts the calibration counter

Factory settings

Off

Calibration counter alarm category

Navigation

Expert ® Sensor ® Diagnostic settings ® Calibration counter alarm category

Description

Use this function to select the category (status signal) as to how the device reacts to the set
calibration countdown.

Options:

• Maintenance required (M)
• Failure (F)

Factory settings

Maintenance required (M)

Calibration counter start value

Navigation

Expert ® Sensor ® Diagnostic settings ® Calibration counter start value

Description

Use this function to set the start value for the calibration counter.

User input

0 to 365 d (days)

Factory settings

365

Calibration countdown

Navigation

Expert ® Sensor ® Diagnostic settings ® Calibration countdown

Description

Use this function to view the time remaining until the next calibration.

The countdown of the calibration counter is only running if the device is switched on.
Example: The calibration counter is set to 365 days on January 1st, 2011. If the device will
be switched off for 100 days, the calibration counter alarm is displayed on April 10th, 2012.