beautypg.com

Ä 52) – Pyromation T82 User Manual

Page 52

background image

Operating menu and parameter description

T82

52

Options:

• Sensor 1 (connection type 1): 2-wire, 3-wire, 4-wire
• Sensor 2 (connection type 2): 2-wire, 3-wire

Factory settings

• Sensor 1 (connection type 1): 4-wire
• Sensor 2 (connection type 2): 3-wire

2-wire compensation n

Navigation

Setup ® 2-wire compensation n
Expert ® Sensor ® Sensor n ® 2-wire compensation n

Prerequisite

An RTD sensor with a 2-wire connection type must be specified as the sensor type.

Description

Use this function to specify the resistance value for two-wire compensation in RTDs.

User input

0 to 30 Ohm

Factory settings

0

Reference junction n

Navigation

Setup ® Reference junction n
Expert ® Sensor ® Sensor n ® Reference junction n

Prerequisite

A thermocouple (TC) sensor must be selected as the sensor type.

Description

Use this function to select reference junction measurement for temperature compensation of
thermocouples (TC).

• If Fixed value is selected, the compensation value is specified via the RJ preset value

parameter.

• Temperature measurement must be configured for channel 2 if Sensor 2 value is

selected.

Options:

• No compensation: no temperature compensation is used.
• Internal measurement: the internal reference junction temperature is used.
• Fixed value: a fixed preset value is used.
• Sensor 2 value: the measured value of sensor 2 is used.

It is not possible to select the Sensor 2 value option for the Reference junction 2
parameter.

Factory settings

Internal measurement

RJ preset value n