Operating menu and parameter description

75

Call./v. Dusen coeff. A, B and C

Navigation

Expert ® Sensor ® Sensor n ® Linearization ® Call./v. Dusen coeff. A, B, C

Prerequisite

The RTD platinum (Callendar/Van Dusen) option is enabled in the Sensor type parameter.

Description

Use this function to set the coefficients for sensor linearization based on the Callendar/Van Dusen
method.

Factory settings

• A: 3.910000e-003
• B: -5.780000e-007
• C: -4.180000e-012

Polynomial coeff. R0

Navigation

Expert ® Sensor ® Sensor n ® Linearization ® Polynomial coeff. R0

Prerequisite

The RTD poly nickel or RTD polynomial copper option is enabled in the Sensor type parameter.

Description

Use this function to set the R0 value for linearization of nickel/copper sensors.

User input

40.000 to 1

050.000 Ohm

Factory settings

100.00 Ohm

Polynomial coeff. A, B

Navigation

Expert ® Sensor ® Sensor n ® Linearization ® Polynomial coeff. A, B

Prerequisite

The RTD poly nickel or RTD copper polynomial option is enabled in the Sensor type parameter.

Description

Use this function to set the coefficients for sensor linearization of copper/nickel resistance
thermometers.

Factory settings

Polynom coeff. A = 5.49630e-003
Polynom coeff. B = 6.75560e-006

"Diagnostic settings" submenu

Calibration counter start

Navigation

Expert ® Sensor ® Diagnostic settings ® Calibration counter start