Ä 58) – Pyromation T82 User Manual
Page 58

Operating menu and parameter description
T82
58
Options:
• Off
• Out band (drift)
• In band
Factory settings
Off
Drift/difference alarm category
Navigation
Setup ® Advanced setup ® Sensor ® Drift/difference alarm category
Expert ® Sensor ® Diagnostic settings ® Drift/difference alarm category
Prerequisite
The Drift/difference mode parameter must be activated with the Out band (drift) or In band
option.
Description
Use this function to select the category (status signal) as to how the device reacts when a drift/
difference is detected between sensor 1 and sensor 2.
Options:
• Out of specification (S)
• Maintenance required (M)
• Failure (F)
Factory settings
Maintenance required (M)
Drift/difference set point
Navigation
Setup ® Advanced setup ® Sensor ® Drift/difference set point
Expert ® Sensor ® Diagnostic settings ® Drift/difference set point
Prerequisite
The Drift/difference mode parameter must be activated with the Out band (drift) or In band
option.
Description
Use this function to configure the maximum permissible measured value deviation between
sensor 1 and sensor 2 which results in drift/difference detection.
Options:
1.0...999.0
Factory settings
999.0
Sensor switch set point
Navigation
Setup ® Advanced setup ® Sensor ® Sensor switch set point
Expert ® Sensor ® Diagnostic settings ® Sensor switch set point
Description
Use this function to set the threshold value for sensor switching (® д 53).