KROHNE IFC 210 E-EEx EN User Manual
Page 9
Part A System installation and start-up Sect. 1.3.2
05/2002
IFC 210 E
9
1.3.2 Stripping (preparation) of signal cables
Please note: The numbers in the drawings designate the stranded drain wires of signalling cables
A and B, see sectional drawings in Sect. 1.3.1.
primary head
Converter
Length primary
head
Length Converter
mm
inch
mm inch
a 90 3.60
a 40 2.80
b 8
0.30
b
10 0.30
c 25 1.00
d
8
0.30
d
5
0.30
e
70 2.80
e 20 2.00
Signal cable A (type DS),
double shielding
for primary head
Signal cable A (type DS),
double shielding
for IFC 210 E Converter
Signal cable B (type BTS),
with triple shielding (bootstrap)
for primary head
Signal cable B (type BTS),
with triple shielding (bootstrap)
for IFC 210 E Converter
Customer-supplied materials
V
Tin-coat all stranded drain wire ends!
W
Insulation tubing (PVC), Ø 2.0-2.5 mm (Ø 1”)
X
Heat-shrinkable tubing or cable sleeve
Y
Wire end sleeve to DIN 41 228: E 1.5-8
Z
Wire end sleeve to DIN 41 228: E 0.5-8
Signal cable B
bending radius
≥
50 mm (
≥
2”)
bending radius
≥
50 mm (
≥
2”)
Signal cable A
bending radius
≥
50 mm (
≥
2”)
bending radius
≥
50 mm (
≥
2”)