KROHNE IFC 210 E-EEx EN User Manual
Page 45
Part B IFC 210 Signal converter
Sect. 5.9
05/2002
IFC 210 E
45
5.9 Controlinput
Controlinput: Fct. 1.07 (output terminal B1) and Fct. 1.08 (output terminal B2)
Press
→
key.
Select function of control inputs,
Select with
↑
or
↓
key.
•
Off
switched off, no function
•
Hold Value
hold value of outputs and display
•
Zero Value
set outputs and display to “min. values”
•
Total Reset
reset totalizers
•
Error Reset
delete/acknowledge error messages
•
Extern Range
external range change
setting range:
05 – 80 %
ratio of high to low range, 1:20 to 1:1,25, (value must be greater
than that of Fct. 1.03 L.F.Cutoff, s. also Sect. 5.3)
Transfer to number setting with
↵
key,
1st number (cursor)with negative image.
Change negative-image number (cursor) with
↑
or
↓
key, move
cursor 1 place to right or left with
→
and
←
key, resp..
Press
↵
key to return to Fct. 1.07 or 1.08 Statusoutput.
Characteristics of status inputs
inactive (no voltage)
active (voltage present)
Off
no function
Extern Range
high range
low range
Hold Value
Measured values follow the
measurement
Hold measured values
Zero Value
Measured values follow the
measurement
Set measured values to
“zero”
Total Reset inactive
Reset
totalizer
Error Reset
inactive
Delete error messages
5.10 F/R mode, forward/reverse flow measurement
•
Refer to Sect. 2.5 for electrical connection of outputs.
•
Define direction of forward (normal) flow, see Fct. 3.02, subfunction Flow Dir. :
in conjunction with F/R operation, set the direction for the forward flow here.
„+“ signifies the same direction as shown by the arrow on the primary head,
„-“ signifies the opposite direction.
•
Set the status output to SIGN I or SIGN P , see Fct. 1.07.
•
Current and/or pulse output must be set to 2 Directions,
see Fct. 1.05 and/or 1.06, subfunction “Function”.