Part B IFC 210 Signal converter

Sect. 5.9

05/2002

IFC 210 E

45

5.9 Controlinput

Controlinput: Fct. 1.07 (output terminal B1) and Fct. 1.08 (output terminal B2)
Press

key.

Select function of control inputs,

Select with

or

key.

Off

switched off, no function

Hold Value

hold value of outputs and display

Zero Value

set outputs and display to “min. values”

Total Reset

reset totalizers

Error Reset

delete/acknowledge error messages

Extern Range

external range change

setting range:

05 – 80 %

ratio of high to low range, 1:20 to 1:1,25, (value must be greater

than that of Fct. 1.03 L.F.Cutoff, s. also Sect. 5.3)

Transfer to number setting with

key,

1st number (cursor)with negative image.

Change negative-image number (cursor) with

or

key, move

cursor 1 place to right or left with

and

key, resp..

Press

key to return to Fct. 1.07 or 1.08 Statusoutput.

Characteristics of status inputs

inactive (no voltage)

active (voltage present)

Off

no function

Extern Range

high range

low range

Hold Value

Measured values follow the
measurement

Hold measured values

Zero Value

Measured values follow the
measurement

Set measured values to
“zero”

Total Reset inactive

Reset

totalizer

Error Reset

inactive

Delete error messages



5.10 F/R mode, forward/reverse flow measurement

Refer to Sect. 2.5 for electrical connection of outputs.

Define direction of forward (normal) flow, see Fct. 3.02, subfunction Flow Dir. :
in conjunction with F/R operation, set the direction for the forward flow here.
+“ signifies the same direction as shown by the arrow on the primary head,
-“ signifies the opposite direction.

Set the status output to SIGN I or SIGN P , see Fct. 1.07.

Current and/or pulse output must be set to 2 Directions,
see Fct. 1.05 and/or 1.06, subfunction “Function”.